Document généré le 30/04/2026 depuis l'adresse: https://www.documentation.eauetbiodiversite.fr/fr/notice/validation-d-un-procede-automatise-d-extraction-sur-phase-solide-de-glandes-digestives-de-moules-pour-l-identification-et-la-quantification-des-dinophysistoxines-en-lc-esi-sm2-par-piegeage-d-ions-quadripolaire
Validation d’un procédé automatisé d’extraction sur phase solide de glandes digestives de moules pour l’identification et la quantification des dinophysistoxines en LC/ESI/SM2 par piégeage d’ions quadripolaire
Titre alternatif
Producteur
Contributeur(s)
Éditeur(s)
Identifiant documentaire
9-45270
Identifiant OAI
oai:archimer.ifremer.fr:45270
Auteur(s):
Mondeguer, Florence,Mestres, Gilberte,Rouland, Caroline,Montagu, Monique,Maillols, Hélène
Mots clés
Toxines
intoxication des coquillages
dinophysistoxines
acide okadaïque
dinophysistoxine-1
méthode alternative
extraction sur phase solide
automatisation
couplage CLHP/ESI/SM2 par piège d’ions.
Toxins
shellfish poisoning
dinophysistoxins
okadaic acid
dinophysistoxin-1
alternative method
solid phase extraction
automation
liquid chromatography coupled ion trap.
Date de publication
01/02/2003
Date de création
Date de modification
Date d'acceptation du document
Date de dépôt légal
Langue
fre
Thème
Type de ressource
Source
Droits de réutilisation
2003 Ifremer
Région
Département
Commune
Description
This report presents a new extraction method of the dinophysistoxins (DTXs), confirmed by quantification using high-performance liquid chromatography coupled to mass spectrometry with an ion trap and electro spray interface (HPLC/ESI/MS2).
The method originality consists on the adaptation of DTXs basic extraction procedure (liquid/ liquid) to a solid phase extraction (SPE) via a robotic station: ASPEC XLi
The parameters of the automatization procedure were optimized to obtain the best DTXs recovery rate.
These improvements were loaded with digestive gland mussel homogenat realized on a silica cartridge SPE, activated in hexane/chloroform (50:50), washed with hexane/chloroform (50:50) and extracted by an elution gradient (chloroform methanol (65:35) and methanol (100%)). This method was validated according to two normative referentials (linearity, detection quantification limits and accuracy…) :
- The Guide of the Pharmacy industry: Analytical Validation, report of the commission SFSTP 1992 (French Corporation of the Sciences and Technical Pharmaceutical),
- - The Procedure of validation of an alternative method in compare to a reference method. (AFNOR, 1998. NF V 03-110).
Comparison with the classical liquid/liquid extraction and the automated method present clear advantages.
In an analytical method the extraction is generally considered to be the most labor-intensive and error-prone step.
This new procedure allowed us to increase throughput, to improve the reproducibility and to reduce the error risks due to the individual manual treatments.